Technical features
Reference Standards and Directives
Directive 93/42/EEC
Quality Ass. standard EN ISO 9001 (2000), UNI EN 13485 (2002)
Safety standards:EN60601-1, EN60601-2-10, EN60601-1-2
Feeding: 230V ~ / 50 Hz
Power consumption: 22VA
Internal fuse: T 500mA
Impulsive current max: 60mA
Continuous current max: 20mA
Output voltage: 100 Vdc
Number of conduits: 2
Treatment time: from 1 to 60 min
Output param. indicator: Alphanumeric Display
Output indicator: Visual
Electric safety class: II BF
Protection level: IP20
Function
– Current Faradic
(Rectangular, Triangular, Exponential)
Galvanic association: Yes
Polarity: Positive / Negative
Impulse duration: 1 μ sec
Frequency: 1, 2, 5, 10, 20, 50, 100, 200, 500Hz
Action time: from 3 to 60 sec
Break time: from 0 to 60 sec
– Current impulsive
(Rectangular, Triangular, Exponential)
Galvanic association: No
Polarity: Positive / Negative
Impulse duration: 1, 2, 5, 10, 20, 50, 100, 200 μsec
Break time: from 1 to 10 sec
– Current diadinamic
Galvanic association: Yes
Polarity: Positive / Negative
MF Frequency: 50Hz
MFS Frequency: 50Hz
Action time: 1 sec
Break time: 1 sec
DF Frequency: 100Hz
DFS Frequency: 100Hz
Action time: 1 sec
Break time: 1 sec
CP: Frequency 50Hz for 1 sec
100Hz for 1 sec
LP: Frequency 50Hz for 5 sec
100Hz for 7 sec
– Current T.E.N.S
Galvanic association: No
Polarity: Positive / Negative
Impulse duration: 100, 150, 200, 250, 300 μsec
Frequency: 1, 2, 5, 10, 20, 50, 100, 200, 500Hz