Technical features
– Input circuit: Floating input circuit, protection from defibrillation and pace maker
– Cable: Standard 12 LEAD
– Acquisition mode: 12 leads simultaneously
– A/D converter: 12 bit
– Measurement range: > ± 5mVpp
– Time constant: > 3.2 sec
– Frequency Response: 0,05Hz-150Hz
– Ecg parameter and analysis: Yes
– Memory: 100 patients
– Calibration voltage: 1mV ± 3%
– Sensitivity: 2,5-5-10-20(mm/mV)
– Input impendance: > 50Ω
– Input Circuit current: 50nA
– Noise level: < 15μ Vp-p
– Anti baseline drift: Auto
– Patient current leakage: < 10μ ( 220V-240V7Hz)
– Patient auxiliary current leakage: < 0,1μ (DC)
– Filters: EMG filter 35Hz ( -3dB) AC filter 50Hz/60Hz (-20dB)
– CMRR: >100 dB
– Recording mode: Auto/manual
– Rhythm lead: Any lead selectable
– Lead selection: Auto/manual
– Safety standard: IEC I/CF
– Power supply: AC:100V-115V/220V-240V,50/60Hz
– Built-in rechargeable Lithion battery: Voltage =14.4V (3hrs continuously operation)
– Printer: Thermal
– Paper: 80mmx30m
– Paper speed 5mm/s, 10mm/s, 25mm/s, 50mms ±3
– Display: 320×240 pixels LCD
– Communication interface: RS232, USB, ETHERNET
– Dimensions: L:28,8xW:21xH:7 cm
– Weight: 2,5 kg