Technical features
– Circuit: protection from defibrillation and pace maker
– Lead: standard 12 leads
– Acquisition mode: 12 leads simultaneously
– A/D converter: 12 bits
– Input Voltage: Range: ±5 Vpp
– Time constant: > 3,2s
– Frequency range: 0,05Hz-150Hz
– Calibration voltage: 1mV± 2%
– Sensitivity: 2,5-5-10,20,10/5 ACG (mm/mV)
– Input Impedance: 50 Mohm (10 Hz)
– Noise level: < 12μ Vp-p
– Anti Baseline Drift: automatic
– Filters: EMG filter, 25Hz/35Hz/ 45Hz/OFF, DFT 0.05/0.15/0.25/0.5 Hz, Low pass filter 150Hz/100Hz/75Hz AC 50 Hz/60Hz CMRR: > 110dB
– Recording mode: Auto/manual/rhythmic
– Rhythm lead: single lead or 3 leads selectable
– Safety: class “I”, “CF” type
– Power supply: AC100V-240V, 50-60Hz
– Battery SE1200: built-in Li-ion, 14.4V
– Printer: integrated thermal printer
– Paper: Z-fold and roller (mm210/216h)
– Display SE1200: 800 x600 single colour LCD
– Memory SE1200: 200 patients
– Input: > 100 K Mohm
– Output: < 100 Mohm
– Communication interface: RJ45(Ethernet), RS232, USB
Dimensions: L:42 x W:33 x H:12 cm
Weight: 9 kg