PATIENT MONITOR MULTIPARAMETER – DISPLAY 10,1″


High resolution colour TFT-LCD display. Visualization of ECG, RESP, SpO2, NIBP, PR, 2-TEMP. Suitable for adult, pediatric and neonatal patients. Arrhythmia analysis. Pacemaker detection, defibrillation and electrosurgical interference protection. 96 hours trend review, 500 NIBP measurements and 60 alarm memory. USB port. With handle for transport. Optional built-in printer.


Accessori INCLUDED


– 1 Five lead ECG Cable, snap, Defib, IEC, 3.5m, reusable

– 30 ECG Electrodes, adult, disposable

– 1 SPO2 Finger Sensor, adult, 2.5m, reusable

– 1 NIBP Cuff, Adult, 25cm-35cm, reusable

– 1 NIBP Tube, 3m

– 1 Temperature Probe Skin, adult, 3m, reusable

– 1 Power Cord (EUR standard)

– 1 Rechargeable Lithium-Ion Battery (14.8V, 2100mAh)

– 1 Ground Cable



Languages available on request 

 Italian

 English

 Spanish

 French

 Portuguese

 German

 Polish

 Chinese

 Turkish

PRODUCT CODE: LTD340 Category

Description

Display 

Colour TFT LCD: 10,1” 

Resolution: 800×480 

Traces displayed: Up to 8 

Waveforms displayed: Up to 11 

Sweep Speed: 6.25, 12.5, 25, 50 mm/s 

ECG 

Lead Type: 5-lead and 3-lead selectable 

Gain: Auto, x 0.125, x 0.25, x 0.5, x 1, x 2, x 4 

Sweep Speed: 6.25, 12.5, 25, 50 mm/s 

ECG HR Range:

Adult: 15~300bpm 

Pediatric/Neonate: 15~350bpm 

Resolution: 1 bpm 

Accuracy: ±1 bpm or ±1% (whichever is greater) 

Filter:  Diagnostic mode 0.05 – 150 Hz / Monitoring mode 0.5 –  40 Hz / Surgical mode 1 – 20 Hz

Diagnostic Mode: 0.05~150 Hz 

Monitoring Mode: 0.5~40 Hz 

Surgical Mode: 1~20 Hz 

ST-Segment Detection:

Measurement Range: -2.0 mV~2.0 mV 

Alarm Range: -2.0 mV~2.0 mV 

RESP 

Method: Trans-thoracic impedance 

Operation Mode: Auto/Manual 

– RR Measurement Range:

Adult:0~120 rpm 

Neonate/Pediatric: 0~150 rpm 

Resolution: 1 rpm 

Apnea Alarm Thresold: 10,15,20,25,30,35,40 s 

Band Width: 0.2-2.5 Hz (-3 dB) 

Sweep Speed: 6.25, 12.5, 25, 50 mm/s 

SPo2 

Measurement & Alarm Range: 0~100% (EDAN SpO2) 

Resolution: 1% 

Accuracy:

±2% (70~100% Adult/Pediatric) 

±3% (70~100% Neonate) 

PR Measurement: 25~300 bpm 

Resolution: 1bpm 

Refresh Rate: 1 second 

NIBP 

Method: Automatic Oscillometric 

Operation Modes: Manual/Automatic/Continuous 

Auto Measurement Time Interval:

1,2,3,4,5,10,15,30,60,90,120,240,480 minutes 

Measurement Unit: mmHg/kPa 

Measurement Types: Systolic, Diastolic, Mean 

Pressure Range:

Adults:

Systolic: 40~270 mmHg 

Diastolic: 10~215 mmHg 

Mean: 20~235 mmHg 

Pediatrics:

Systolic: 40~200 mmHg 

Diastolic: 10~150 mmHg 

Mean: 20~165 mmHg 

Neonates:

Systolic: 40~135 mmHg 

Diastolic: 10~100 mmHg 

Mean: 20~110 mmHg 

Resolution: 1 mmHg 

Accuracy:

Max Mean error: ±5 mmHg 

Max Standard Deviation: 8 mmHg 

PR from NIBP Measurement Range: 40~240 bpm 

Resolution: 1bpm 

Accuracy: 3 bpm or 3.5% (whichever is greater) 

SP10: 2002 

Temperature (2 Channels, 1 probe by default) 

Measurement/Alarm Range: 0~50C° (32-122 F°) 

Resolution: 0.1 C° 

Accuracy: ±0.1 C° (without probe) 

Channel: Dual-Channel- Provide T1; T2; ΔT 

Power Supply

AC Power: 100~240 V AC, 50/60 Hz 

Battery: 14.8 V Rechargeable Li-ion Battery

Dimensions:  L:32 x W:15 x H:26,5 cm

Weight: 5 kg

Close
Close