VITAL SIGNS MONITORS M3A SpO2 + NIBP, 2 displays


Created to be used in hospital and at home. 72 hours internal memory. 5 hours for SpO2 wave form, 2,5 hours for SpO2+CO2 wave form. 30.000 NIBP measurements and 800 alarms can be visualized.


 


Optional printer.


 


2 displays 


– 3,5” LCD colour display


– LED colour display 


 


 

PRODUCT CODE: LTD425 Category

Description


Dimension and weight


L: 174 mm W: 189 mm H: 235 mm


3,5 kg (without batteria)


 








































Specification 


Anti-electroshock degree 


SpO2, NIBP, BF Defibrillation type 


Ingress Protection 


IPX1 (No protection against ingress

of water if configured with TEMP module) 


WORKING CONDITIONS 


Working Temperature 


5 ~ 40 °C 


Power Supply 


100 ~ 240 VAC, 50/60Hz, 


Pmax = 45VA FUSE T 1.6AL 


Display 


3,5” LCD colour display


Resolution: 640×480 


BATTERY 


Type 


1 Li battery 


Power-off delay 


5 ~ 15min (After the low battery alarm) 


Voltage 


14.8 VDC 


Working period 


± 240 min 


Rechargeable period 


± 180 min 


PRINTER 


Record wdth 


48 mm 


Paper Speed 


25mm/s 


Recording type 


Current displayed parameter list recording 


Current displayed alarm list recording 


Trend List Recall 


72 hours, 1 minute Resolution 


NIBP Measurement Recall 


30, 000 NIBP measurement data 


NIBP 


Method 


Oscillometric 


Mode 


Manual, Auto, Continuous 


MEASURING INTERVAL 


Measuring interval 


1/2/3/4/5/10/15/30/60/90/120/240/480 min 


Continuous 


5min, interval is 5s 


Measuring Type 


Systolic Pressure,

Diastolic Pressure,

Mean Pressure 


MEASURING RANGE 


Adult Mode 


Pediatric Mode 


SYS 40~270mmHg 


DIA 10~215mmHg 


MAP 20~235mmHg 


SYS 40~200mmHg 


DIA 10~150mmHg 


MAP 20~165mmHg 


CUFF PRESSURE 


Measuring Range 


0~300mmHg 


Pressure Resolution 


± 1 mmHg 


Accuracy 


± 3 mmHg 


Maximum mean error 


± 5 mmHg 


Maximum Standard deviation 


± 8 mmHg 


Typical Measuring Period 


30~45s (depend on HR/motion disturbance)
















Overpressure 


protection 


Dual overpressure protection 


Adult 297±3mmHg 


Pediatric 240±3mmHg 


Neonatal 147±3mmHg 


SpO


Measuring Range 


0 ~ 100 % 


Alarm Range 


0 ~ 100 % 


Resolution 


1 % 


Accuracy 


Adult (including Pediatric) 


± 2 % (70%~100% SpO2


Undefined (0~69% SpO2


Neonate 


± 3 % digits (70%~100% SpO2) 


Undefined (0~69% SpO2) 


PULSE RATE 


Measuring 


25 ~ 300 bpm 


Alarm Range 


30 ~ 300 bpm 


Resolution 


1 bpm 


Accuracy 


± 3 bpm under motion condition 


± 5 bpm 


Anti-motion Interference 


Strong Anti-motion Interference, 


Anti-electrotome 


 

Close
Close